High-resolution Blood Flow Velocity Measurements In The Human Finger

Full text links CiteDisplay options Display options Format AbstractPubMedPMID

Abstract

MR phase contrast blood flow velocity measurements in the human index finger were performed with triggered, nontriggered, and cine acquisition schemes. A strong (G(max) = 200 mT/m), small bore (inner diameter 12 cm) gradient system inserted in a whole body 3 Tesla MR scanner allowed high-resolution imaging at short echo times, which decreases partial volume effects and flow artifacts. Arterial blood flow velocities ranging from 4.9-19 cm/sec were measured, while venous blood flow was significantly slower at 1.5-7.1 cm/sec. Taking into account the corresponding vessel diameters ranging from 800 microm to 1.8 mm, blood flow rates of 3.0-26 ml/min in arteries and 1.2-4.8 ml/min in veins are obtained. The results were compared to ultrasound measurements, resulting in comparable blood flow velocities in the same subjects. Magn Reson Med 45:716-719, 2001.

PubMed Disclaimer

Publication types

  • Comparative Study Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search
  • Research Support, Non-U.S. Gov't Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search

MeSH terms

  • Blood Flow Velocity / physiology* Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search
  • Fingers / blood supply* Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search
  • Humans Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search
  • Magnetic Resonance Imaging / methods* Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search
  • Ultrasonography Actions
    • Search in PubMed
    • Search in MeSH
    • Add to Search

LinkOut - more resources

  • Full Text Sources

    • Wiley
  • Other Literature Sources

    • The Lens - Patent Citations Database
  • Medical

    • MedlinePlus Health Information

Tag » How Fast Does Blood Circulate