Modified HPLC-electrospray Ionization/mass Spectrometry Method ...
Có thể bạn quan tâm
Background: Monitoring of hemoglobin A(1c) (HbA(1c)) is important in the management of diabetes. The IFCC reference measurement procedure for HbA(1c) is based on the ratio of glycated to nonglycated N-terminal hexapeptides of the beta-chains of hemoglobin after digestion with Glu-C endoproteinase. We developed a modification of the original reference measurement procedure with HPLC-electrospray ionization/mass spectrometry (ESI/MS).
Method: We performed chromatographic separation of the hexapeptides using a C12 reversed-phase column and a binary gradient system consisting of a mixture of H(2)O/acetonitrile/formic acid.
Results: Using this method, we obtained higher signal intensities and improved system stability compared with the reference measurement procedure. In the range of 3% to 14% HbA(1c), intralaboratory CVs were 0.71% to 1.86%. Deviations from IFCC target values were -0.87 to 1.00 relative %. These values fulfill acceptability criteria for HbA(1c) determination set by the IFCC Working Group on HbA(1c) Standardization.
Conclusions: This procedure for the determination of HbA(1c) improves the existing reference measurement procedure.
Từ khóa » Hba1c Nach Ifcc (hplc)
-
[PDF] HbA1c Nach IFCC In SI-Einheit - Labor Dr. Berghold
-
A History Of HbA 1c Through Clinical Chemistry And Laboratory ...
-
HbA1c (IFCC)
-
HbA1c: Was Ihr Laborwert Bedeutet
-
HbA1c - Öquasta
-
Hämoglobin-A1c (HbA1c) - Medizinisches Labor Ostsachsen
-
Basiswissen Diabetes: - Der HbA1c-Wert
-
[PDF] HbA1c – Methodenumstellung Und Umstellung Auf Die Neuen IFCC ...
-
HbA1c: Neuer Referenzstandard - Österreichische Ärztezeitung
-
HbA1c: EQA In Germany, Belgium And The Netherlands Using Fresh ...
-
Laborwerte: HbA1c | Apotheken Umschau
-
[PDF] HbA1c FS* - DiaSys Diagnostic Systems
-
Bei Der HbA 1c -Messung Gibt Es Viele Fehlerquellen - SpringerLink