Optimizing Xevo G2-XS QTof Quadrupole Settings To Increase ...

Optimizing Xevo G2-XS QTof Quadrupole Settings to Increase Sensitivity and Dynamic Range for the Analysis of TrastuzumabTechnical notes | 2018 | WatersInstrumentationLC/TOF, LC/HRMS, LC/MS, LC/MS/MSIndustriesPharma & BiopharmaManufacturerWatersGo to the libraryView PDF

Summary

Significance of the Topic

The accurate quantification of protein therapeutics in complex biological matrices is critical for drug discovery, development, and quality control. High-resolution LC-MS methods offer enhanced selectivity and sensitivity compared to traditional ligand binding assays, especially when measuring surrogate peptides in digested samples that contain numerous endogenous interferences.

Objectives and Study Overview

This study aimed to optimize the quadrupole mass selection window on a Xevo G2-XS QTof high-resolution mass spectrometer to maximize sensitivity and dynamic range for two trastuzumab surrogate peptides in rat plasma. By evaluating transmission windows from 1 to 4 Da, the experiment assessed impacts on signal-to-noise ratio, interference removal, and calibration performance.

Methodology and Instrumentation

Sample preparation included affinity capture of trastuzumab from rat plasma, enzymatic digestion using the ProteinWorks eXpress Digest Kit, and peptide cleanup via µElution SPE. Chromatographic separation was performed on an ACQUITY UPLC H-Class System with a Peptide BEH C18 column under an eight-minute gradient (5–50% acetonitrile with 0.1% formic acid, 0.3 mL/min). Mass spectrometry was conducted in ESI+ mode using Time-of-Flight Multiple Reaction Monitoring (Tof-MRM) on the Xevo G2-XS QTof, acquiring m/z 200–1500 and applying quadrupole windows of 1, 3, and 4 Da. Data processing and quantitation were carried out with TargetLynx software using a 50 mDa mass extraction window.

Used Instrumentation

  • Xevo G2-XS QTof high-resolution mass spectrometer (Waters Corporation)
  • ACQUITY UPLC H-Class System and Peptide BEH C18 column (2.1×50 mm, 1.7 µm)
  • ProteinWorks eXpress Digest Kit and µElution SPE Clean-up Kit
  • MassLynx v4.1 and TargetLynx Application Manager

Results and Discussion

Narrowing the quadrupole transmission window to 1 Da sharply reduced co-eluting background signals and isobaric interferences while retaining sufficient precursor ion transmission. Compared with a 4 Da window, the 1 Da setting yielded a 4- to 8-fold increase in signal-to-noise ratio at the lower limit of quantitation for both DTYIHWVR (0.05 µg/mL) and FTISADTSK (0.025 µg/mL). Although absolute peak intensities decreased, interference removal led to more reliable automatic integration. Calibration curves demonstrated that both peptides achieved an extended dynamic range by approximately half an order of magnitude when using the narrow window.

Benefits and Practical Applications

The optimized quadrupole settings facilitate sensitive, selective quantitation of low-abundance peptides in complex matrices, improving assay robustness and throughput. This approach is readily applicable to bioanalytical assays for monoclonal antibodies and other large protein drugs in pharmaceutical research and routine QC laboratories.

Future Trends and Potential Applications

Advances may include adaptive or variable quadrupole windows driven by real-time data analysis and machine learning to further balance sensitivity and selectivity. The strategy can be extended to intact protein quantification, multiplexed peptide panels, and other HRMS platforms, broadening its impact in biopharma and clinical diagnostics.

Conclusion

Optimizing the quadrupole mass selection window on the Xevo G2-XS QTof to 1 Da significantly enhances selectivity and sensitivity for surrogate peptide quantitation in plasma, extends dynamic range, and simplifies data processing. This targeted MS approach supports robust bioanalysis of biotherapeutics.

References

  1. Li F, Fast D, Michael S. Absolute quantitation of protein therapeutics in biological matrices by enzymatic digestion and LC-MS. Bioanalysis. 2011;3(21):2459–2480.
  2. Olah TV. Quantitation of therapeutic proteins following direct trypsin digestion of dried blood spot samples by LC-MS in drug discovery. Bioanalysis. 2012;4(1):29–40.
  3. Lame M, Dunning C, Alelyunas Y, Wrona M. Comparison of tandem and high resolution mass spectrometry for quantification of trastuzumab in plasma. Waters Corporation; 2018. Tech Brief 720006207EN.
  4. Promega Corporation. Magne Protein A beads. Accessed December 2018.
Read morearrow iconDownloadable PDF for viewing 

Similar PDF

ToggleComparison of Tandem and High Resolution Mass Spectrometry for the Quantification of the Monoclonal Antibody, Trastuzumab in PlasmaComparison of Tandem and High Resolution Mass Spectrometry for the Quantification of the Monoclonal Antibody, Trastuzumab in Plasma2018|Waters|Technical notes[ TECHNOLOGY BRIEF ] Comparison of Tandem and High Resolution Mass Spectrometry for the Quantification of the Monoclonal Antibody, Trastuzumab in Plasma Mary Lame, Caitlin Dunning, Yun Alelyunas, and Mark Wrona Waters Corporation, Milford, MA, USA Demand for bioanalytical quantification…
Key words
ftisadtsk, ftisadtskdtyihwvr, dtyihwvrtrastuzumab, trastuzumabtandem, tandemquadrupole, quadrupolexevo, xevopeptide, peptidebrief, briefmrm, mrmbioanalytical, bioanalyticalweighting, weightinglod, lodmass, massperformance, performancetechnologyLC-MS Quantification of Trastuzumab Using Pellet Digestion and ProteinWorks eXpress Direct Digest KitsLC-MS Quantification of Trastuzumab Using Pellet Digestion and ProteinWorks eXpress Direct Digest Kits2016|Waters|ApplicationsLC-MS Quantification of Trastuzumab Using Pellet Digestion and ProteinWorks eXpress Direct Digest Kits Mary Lame, Paula Orens, Sherri Naughton, and Erin Chambers Waters Corporation, Milford, MA, USA G OA L To demonstrate the compatibility of ProteinWorks™ Kits with pellet digestion…
Key words
ppt, pptarea, areaalbumin, albumintrastuzumab, trastuzumabdigestion, digestionpellet, pelletproteinworks, proteinworksconc, concpeptide, peptidepeptides, peptidesexpress, expressdtyihwvr, dtyihwvrdirect, directtryptic, trypticftisadtskWaters Application Notes - ProteinWorksWaters Application Notes - ProteinWorks2016|Waters|GuidesWaters Application Notes ProteinWorks OUR SCIENTISTS Erin E. Chambers As a principal scientist, Erin has been working almost exclusively on peptide and protein bioanalysis for the last seven years, while managing small and large molecule bioanalysis and clinical research applications…
Key words
proteinworks, proteinworksdigest, digestexpress, expressarea, areapeptide, peptidekit, kittrastuzumab, trastuzumabgeneric, genericplasma, plasmaftfsldtsk, ftfsldtskdigestion, digestionmurine, murineantibody, antibodyprotein, proteindstyslsstltlskPEPTIDE AND PROTEIN BIOANALYSISPEPTIDE AND PROTEIN BIOANALYSIS2016|Waters|Guides[ APPLICATION NOTEBOOK ] PEPTIDE AND PROTEIN BIOANALYSIS [ OUR SCIENTISTS ] Yun Alelyunas, PhD Before coming to Waters in 2012, Yun Alelyunas was a principal scientist and team leader at AstraZeneca for 20 years where she was involved in…
Key words
ionkey, ionkeypeptide, peptideplasma, plasmaquantification, quantificationhuman, humanarea, areaxevo, xevoinsulin, insulinuplc, uplcprotein, proteinmrm, mrmpeptides, peptidesproteinworks, proteinworksantibody, antibodyusingOther projectsGCMSICPMSFollow usMore informationWebinarsAbout usContact usTerms of useLabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Từ khóa » Xevo G2-xs