X-ray Photoelectron Spectroscopy (XPS)
The operating principle of XPS is the photoelectric effect which occurs when an electron is ejected from an element after being exposed to a photon whose energy exceeds a critical value. The ejected electron escapes with a kinetic energy equal to the difference between the incident energy of the photon and its original binding energy (or the energy required to remove the electron from its orbital into space). The relation is expressed by the following equation: EK=hn-EB.
An XPS spectrophotometer operates by striking the sample with a high energy photon (an X-ray from a Mg or Al based source) and measuring the kinetic energy of all of the ejected electrons. The binding energies of these electrons are then calculated and from these a distribution of atomic species is determined. The lightest element that XPS can detect is lithium. Even though the X-ray photons can penetrate deeply into the material, scattering limits only electrons from the top 5-10 nm to be ejected without attenuation of their kinetic energy. Thus, XPS is considered a surface sensitive technique.
In addition to determining the atomic composition of the surface, XPS can also provide semi-quantitative information (expressed in units of atomic percent) by dividing the peak areas of each element by sensitivity factor, which is simply the probability of a particular element to emit an electron when struck by a photon. The sensitivity of XPS is element dependent, with larger elements (which have larger sensitivity factors) being more sensitive than smaller elements.
XPS Facilities at the CMSC
A Perkin Elmer Phi 5600 ESCA system is used for XPS analysis with a magnesium Ka X-ray source at a take off angle of 45°. Sample submitted for analysis must be solid and vacuum compatible (i.e. samples should not have a low sublimation pressure or be excessively oily, etc.). The maximum sample dimensions are roughly 2 cm in diameter and 1 cm in height. A pump down time is required to bring the sample to a proper vacuum level.
Location: Room 1130 EngineeringOn This Site
- About
- Facilities & Equipment
- Gallery
- News & Announcements
- Personnel
- Research
- Testing & Analysis
Departments
- Applied Engineering Sciences
- Biomedical Engineering
- Biosystems & Agricultural Engineering
- Chemical Engineering & Materials Science
- Civil & Environmental Engineering
- Computational Mathematics, Science and Engineering
- Computer Science & Engineering
- Electrical & Computer Engineering
- Mechanical Engineering
Contact Us
Engineering Building428 S. Shaw Lane, Room 1415East Lansing, MI 48824Phone: (517) 353-5466
Developed by DECS | Contact Information | Privacy Statement | Site Accessibility Call MSU: (517) 355-6616 | Visit msu.edu | MSU is an affirmative-action, equal-opportunity employer. SPARTANS WILL | © Michigan State University Từ khóa » Phi 5600
-
Refurbished PHI 5600 XPS System W - RBD Instruments, Inc.
-
[PDF] Phi-5600-specifications.pdf - RBD Instruments, Inc.
-
PHI 5600 - National Nanotechnology Coordinated Infrastructure
-
X-ray Photoelectron Spectroscopy (xps)(PHI 5600-ci Spectrometer)
-
Phi 5600 ESCA System - MEMS Exchange
-
Used PHYSICAL ELECTRONICS / PHI 5600 XPS #9035452 For Sale
-
PHYSICAL ELECTRONICS / PHI PHI 5500 / PHI 5600 ... - CAE Online
-
PHI 5600 XPS System Bakeout Procedure To Improve Base Vacuum
-
X-Ray Photoelectron Spectroscopy (XPS) Surface Analysis Technique
-
SOP Starting XPS Electronics In PHI 5600 - YouTube
-
By Camerafik | PHI 5600 ESCA System - Facebook
-
5600 Đô La Mỹ đến Đơn Vị Tiền Tệ Tiêu Chuẩn được Sử Dụng ở Nam Phi